Prof.
Götz Eckold
(Georg-August University of Göttingen)
03/04/2017, 14:30
An analytical approach for the calculation of the 4D-resolution function of a three-axes spectrometer is presented that takes into account all essential features of state-of-the art instruments. In particular, the performance of focusing techniques in direct and reciprocal space is considered in detail. Complementary to existing numerical simulation methods, the analytical treatment provides...
Dr
Jiri Kulda
(Institut Laue-Langevin)
03/04/2017, 13:50
The flexibility of the TAS instrument together with the selective nature of the corresponding experiments (as opposed to a global data collection by the TOF methods) calls for a certain refinement of the experiment strategy prior to the start of the measurements. In particular for the less experienced users, to assess the advantages and disadvatages of diverse optional instrument setups and to...