We are going to present the upgrades developed at NREX during the shutdown time. We installed new sample environment to investigate thin film samples in H2 or D2 atmospheres: A gas handling system with remote pressure control (0-1bar), a sample chamber with both neutron and x-ray windows, and a He3 cryostat (0.5 - 300K) also with neutron and x-ray windows for simultaneous reflectometry measurements. The cryostat fits into the NREX sample electromagnet (0.5T). We developed new sample holders with spring loaded electrical contacts, which permit simultaneous neutron and x-ray reflectometry. These contacts also work at low temperatures without the need to wire-bond the samples. The contact arrangement is of van-der-Pauw type (4 contacts on a square) connected to a relais matrix, such that the current flow and voltage measurement can be freely assigned, either parallel or perpendicular to the external H field, or in crossed geometry for Hall measurements. The voltage is measured by a sensitive lock-in amplifier, and thus very low excitation currents avoiding sample heating are possible.